Neopure’s ADP saturated steam autoclaves are designed for the sterilization of solids (instruments, rubber materials, machinery parts, empty vials, etc.) and porous loads (textiles, filters, etc.).
The autoclave unit features a control system that enables fully automated, reliable, and reproducible sterilization cycles. This is ensured by the high quality of the materials and components used, the calibration of critical instruments, and the implementation of proven safety and alarm systems. The machine is also equipped with instruments to record events and cycle parameters.
The autoclaves undergo comprehensive testing at our factory under simulated final operating conditions, and their compact design allows for easy installation and commissioning at the destination site.
Steam autoclaves are designed for the moist-heat sterilization of solid loads—whether porous or non-porous—and small-volume sealed liquids (ampoules, vials).
Chamber: AISI-316Ti
Chamber body: AISI-316L
Piping, valves, and instrumentation: AISI-316L
Panels: AISI-304
Base frame: AISI-304
The autoclave body consists of an inner chamber and an outer jacket. The inner chamber is constructed entirely of AISI 316 Ti stainless steel with a thickness of 6 mm.
It has a rectangular cross-section, with the four longitudinal edges rounded to a radius of 50 mm to facilitate cleaning.
The jacket features flared bosses designed to serve two purposes: creating the necessary spacing for smooth steam circulation around the inner chamber, and providing the jacket with the structural strength to withstand steam pressure while simultaneously reinforcing the inner chamber.
The jacket is made of 4 mm thick AISI 316 L stainless steel and fully encloses the inner chamber. The interior of the inner chamber is polished to a surface roughness (Ra) of less than 0.64 µm.
The jacket is covered with 70 mm thick mineral wool insulation and clad externally with AISI-304 stainless steel sheeting featuring a satin-polished finish.
The inner chamber is equipped with a validation port for inserting validation probes and for use as a connection point for vacuum and pressure tests. Two guide rails can be installed at the base of the chamber to allow loading carts to slide in, if the autoclave is equipped with them.
Non-proprietary components (Gemü, Siemens, Festo, SMC)
Automated, reliable and reproducible systems designed for the sterilization of instruments, components, textiles, filters and other materials used in pharmaceutical applications.
Would you like more information?
Tell us what you need and our team will get in touch with you.